The TA1300 is a series of high-stability, high-accuracy AC current standard sources. It is designed for calibrating AC ammeters, current transformers, current sensors, and other AC measurement equipment. With precise output and low distortion, the TA1300 ensures reliable performance in laboratories and industrial applications.
AC Current Output
| Range | Short-term Stability (%/min) | Measurement Uncertainty (k=2) (ppm*RD+ppm*RG) | Compliance Voltage (rms) | Degree of Distortion (%) |
|---|---|---|---|---|
| 100 A | 0.005 | Class 0.02: 120 + 180 Class 0.05: 300 + 200 |
0.7 V | <0.5 |
| 300 A | 0.005 | Class 0.02: 120 + 180 Class 0.05: 300 + 200 |
0.7 V | <0.5 |
| 1 kA | 0.005 | Class 0.02: 120 + 180 Class 0.05: 300 + 200 |
0.7 V | <0.5 |
| N*300 A | 0.005 | Class 0.02: 120 + 180 Class 0.05: 300 + 200 |
0.7 V | <0.5 |
General Specifications
| Specification | Details |
|---|---|
| Power Supply | AC (220 ± 22) V, (50 ± 2) Hz |
| Maximum Power Consumption | 1 kA: 2 kVA 2 kA: 4 kVA |
| Warm-up Time | 30 minutes |
| Temperature Performance | Working temperature: 0°C ~ 40°C Calibration temperature: 18°C ~ 28°C Storage temperature: -20°C ~ 70°C |
| Humidity Performance | Operating humidity: <80% @ 30°C, <70% @ 40°C Storage humidity: 20%~80% R·H, no condensation |
| Altitude | < 3000 m |
| Communication Interface | RS232 interface |
Key Features
- High-stability, high-accuracy AC current output.
- Wide output range: 100 A up to N*300 A.
- Low distortion (<0.5%) ensures precise calibration.
- Short-term stability: 0.005%/min for Class 0.02 and 0.01%/min for Class 0.05.
- Low uncertainty (ppm*RD + ppm*RG) across wide current ranges.
- Supports calibration of AC ammeters, current transformers, and sensors.
- Robust design for laboratory and industrial use.