TD2550 Non-touch Static Voltmeters Calibration Device
TD2550 is a high-precision and high-stability DC high-voltage standard source designed for applications requiring extremely accurate and repeatable high-voltage output. With a maximum output of 20 kV (extendable to 30 kV upon request), it is ideal for calibration of non-contact electrostatic voltmeters, high-voltage electrostatic meters, DC high-voltage measurement systems, and high-voltage probes. Its advanced control technology ensures excellent stability, low drift, and high accuracy suitable for metrology laboratories, power utilities, automation industries, and calibration facilities.
High Voltage Output Specifications
| Range | Resolution | Stability (%/min) | Accuracy (± ppm of reading + ppm of range) | Max Burden (mA) |
|---|---|---|---|---|
| 5 kV | 10 mV | 0.03 (Class 0.1) 0.02 (Class 0.05) |
Class 0.1: 600 + 400 Class 0.05: 300 + 200 |
1.5 |
| 20 kV | 100 mV | 0.03 (Class 0.1) 0.02 (Class 0.05) |
Class 0.1: 600 + 400 Class 0.05: 300 + 200 |
1.5 |
| 30 kV[2] | 100 mV | 0.03 (Class 0.1) 0.02 (Class 0.05) |
Class 0.1: 600 + 400 Class 0.05: 300 + 200 |
1.0 |
General Specifications
| Power Supply | AC (220 ± 22)V, (50 ± 2)Hz |
|---|---|
| Operating Temperature | 0â ~ 45â |
| Storage Temperature | -20â ~ 70â |
| Operating Humidity | < 80% @ 30â, < 70% @ 40â, < 40% @ 50â |
| Storage Humidity | 20% ~ 80% RH (Non-condensing) |
| Communication Interface | RS232 |