UC9908S Hipot Tester
UC9908S Hipot Tester

UC9908S Hipot Tester

Models:
UC9908S (Host, 8-channel) | UC1908S (Slave, 8-channel)

The UC9908S series is a high-performance programmable safety tester with AC/DC withstand voltage testing, insulation resistance (IR) testing, scan test, and contact checking functions. Optional expansion allows up to 88-channel testing. It ensures safe and accurate testing of instruments, equipment, and components.

Performance Features

  • Host device: UC9908S, Slave device: UC1908S
  • Test Functions: ACV / DCV / IR / SCAN / Contact Check
  • Large LCD display for easy reading
  • AC Voltage Output: 50V–5kV, AC Current: 30mA
  • DC Voltage Output: 50V–6kV, DC Current: 10mA
  • Wide insulation resistance measurement range: 1MΩ–50GΩ
  • Optional Ground Connection Test (GC)
  • Programmable leakage current limits
  • Rapid IR/DC discharge function
  • Standard RS-232 communication; optional GPIB interface

Technical Specifications

Model Function Output Channels AC Output DC Output Insulation Resistance (IR) Output Current
UC9908S (Host) ACV/DCV/IR/SCAN/CK 8 channels 0.05–5kV, ±1% ±5V, Display: 2V, 50/60Hz, Max Power: 150VA 0.05–6kV, ±1% ±5V, Display: 2V, Max Power: 60VA 0.05–1kV, 1MΩ–50GΩ, Display accuracy: 0.1MΩ AC: 30mA, ±1% ±5 counts, Display: 1µA
DC: 10mA, ±1% ±5 counts, Display: 0.1µA
UC1908S (Slave) ACV/DCV/IR/SCAN 8 channels 0.05–5kV, ±1% ±5V 0.05–6kV, ±1% ±5V 0.05–1kV, 1MΩ–10GΩ, Display accuracy: 0.1MΩ AC: 30mA, DC: 10mA

ARC Test & Protection

  • Programmable ARC detection
  • Cut-off current: AC 1–15mA, DC 1–10mA
  • Detection time: ~10µs
  • Fast cut-off time: 0.4ms after NG event
  • Fast discharge time: 0.2s
  • Ground failure starting current: 0.5mA ±0.25mA AC

Dimensions & Weight

  • Carton Package: 470 × 280 × 550 mm
  • Weight: 17 kg

Accessories

  • UC90001R High voltage test cable (each channel)
  • UC90001B Ground cable
  • Power cord